Publication
Avaliação da qualidade e eficácia dos fotopolimerizadores na Clínica Universitária da Universidade Católica Portuguesa e em consultórios privados de medicina dentária
datacite.subject.fos | Ciências Médicas::Ciências da Saúde | pt_PT |
dc.contributor.advisor | Noites, Rita Brandão de Pinho | |
dc.contributor.advisor | Almeida, Carlos Emanuel Natividade Ferreira de | |
dc.contributor.author | Braga, Rodrigo João Picado | |
dc.date.accessioned | 2019-10-21T15:02:52Z | |
dc.date.available | 2019-10-21T15:02:52Z | |
dc.date.issued | 2019-07-24 | |
dc.description.abstract | Introdução: Numa era da medicina dentária baseada cada vez mais na “adesão” e com materiais fotopolimerizavéis é fundamental que os fotopolimerizadores emitam luz com intensidade superior a 300mW/cm2. Tivemos como objetivo avaliar o desempenho dos aparelhos fotopolimerizadores da Clínica Universitária da UCP – Viseu bem como em clínicas privadas de Viseu. Materiais e Métodos: Recorrendo a dois radiómetros digitais, foram realizados testes na clínica Universitária da UCP bem como em clínicas privadas com valência em medicina dentária. Avaliou-se a radiância debitada por cada fotopolimerizador em mili-watts por centímetro quadrado (mW/cm2). Utilizou-se um radiómetro SDI Led Radiometer (SDI, Austrália) e um radiómetro Led Radiometer (Woodpecker, China). No caso de fotopolimerizadores de quartzo-tungsténio-halogénio (QTH) foram realizadas três exposições prévias de 60 segundos consecutivas com intervalo de 1 segundo de descanso. Para as unidades LED foram efetuadas três leituras para cada aparelho. Para cada aparelho foram testados os tempos de polimerização de 20 segundos. Resultados: Foram avaliados 18 aparelhos na Clínica Universitária da UCP – Viseu, sendo 50% de QTH e 50% de LED. 23.5% dos fotopolimerizadores possuía danos e 42% tinha presença de resíduos nas suas pontas. O valor de intensidade média da luz emitida foi de 175 mW/cm2 nos fotopolimerizadores Quartzo-Tungsténio-Halogénio e 1190 mW/cm2 para os LED (p<0.005). Nas clínicas privadas foram analisados 30 fotopolimerizadores LED, dos quais 36,7% apresentava danos na ponta e 26,7% presença de resíduos. O valor médio de intensidade de luz obtido foi de 1170mW/cm2 (p<0.005). Conclusão: Conclui-se que na Clínica Universitária, apenas unidades de LED devem ser usadas. Os resultados das unidades de LED permitem que a fotopolimerização ocorra dentro dos parâmetros internacionalmente aceites. Nas clínicas privadas todos os fotopolimerizadores possuíam intensidades de fotopolimerização muito acima do recomendado pelos fabricantes pelo que é necessário adotar o modo progressivo de polimerização de forma a reduzir o stress de polimerização induzido. | pt_PT |
dc.description.abstract | Introduction: In an era of dental medicine based on adhesion materials, it is fundamental to ensure the light best conditions and be able to produce 300mW/cm2. We aimed to evaluate the performance and conservation of the light curing devices of the UCP - Viseu University Clinic. Material and Methods: Two digital radiometers were used, an SDI Led Radiometer (SDI, Australia) and a Led Radiometer (Woodpecker, China) to evaluate light curing units from UCP University Clinic and private dental offices. The radiance charged by each light curing unit in milli-watts per square centimeter (mW/cm2) was evaluated. In the case of quartz-tungsten-halogen (QTH) light curing units were performed three previous exposures of 60 consecutive seconds with a rest interval of 1 second. For the LED devices- three reading was done for each unit. For each unit, a 20-second exposure was tested. For all the tests, p≤0,05 was used statistically. Results: Eighteen devices were evaluated at UCP University Clinic: 50% QTH and 50% LED at UCP - Viseu University Clinic. Damage was present in 23.5% of the analyzed units, and the presence of residues in its tip occurs in about 42%. The intensity value of the emitted light was 175 mW/cm2 for the QTH and 1190 mW/cm2 for the LEDs (p <0.005). In private dental offices, 30 LED light curing units were evaluated. 36,7% presented tip damage and 26,7% had residues in its tip. The intensity value of the emitted light was 1170 mW/cm2 (p <0.005). Conclusion: With this study, at the UCP – Viseu University Clinic only LED units should be used. The results of the LED units allow the photopolymerization to take place within the internationally accepted parameters. Private dental offices emitted light shown too higher values so is advisable to use the progressive mode in order to decrease the top layer polymerization stress. | pt_PT |
dc.identifier.tid | 202280608 | pt_PT |
dc.identifier.uri | http://hdl.handle.net/10400.14/28458 | |
dc.language.iso | por | pt_PT |
dc.subject | Medicina dentária | pt_PT |
dc.subject | Fotopolimerização de adesivos | pt_PT |
dc.subject | Fotopolimerizadores | pt_PT |
dc.subject | Resinas compostas | pt_PT |
dc.subject | Polimerização | pt_PT |
dc.subject | Dentistry | pt_PT |
dc.subject | Light curing of dental adhesives | pt_PT |
dc.subject | Curing lights | pt_PT |
dc.subject | Composite resins | pt_PT |
dc.subject | Polymerization | pt_PT |
dc.title | Avaliação da qualidade e eficácia dos fotopolimerizadores na Clínica Universitária da Universidade Católica Portuguesa e em consultórios privados de medicina dentária | pt_PT |
dc.type | master thesis | |
dspace.entity.type | Publication | |
rcaap.rights | openAccess | pt_PT |
rcaap.type | masterThesis | pt_PT |
thesis.degree.name | Mestrado em Medicina Dentária | pt_PT |
Files
Original bundle
1 - 1 of 1
No Thumbnail Available
- Name:
- Dissertação Rodrigo Braga 500114008.pdf
- Size:
- 915.66 KB
- Format:
- Adobe Portable Document Format
License bundle
1 - 1 of 1
No Thumbnail Available
- Name:
- license.txt
- Size:
- 3.44 KB
- Format:
- Item-specific license agreed upon to submission
- Description: