Use of a rhodamine-based chelator in a microfluidic paper-based analytical device for the in-situ copper quantification in natural waters

dc.contributor.authorAguiar, Juliana I. S.
dc.contributor.authorRibeiro, Susana O.
dc.contributor.authorLeite, Andreia
dc.contributor.authorRangel, Maria
dc.contributor.authorRangel, António O. S. S.
dc.contributor.authorMesquita, Raquel B. R.
dc.date.accessioned2024-02-12T11:39:43Z
dc.date.available2024-02-12T11:39:43Z
dc.date.issued2024-05-01
dc.description.abstractThis work describes the development of a microfluidic paper-based analytical device (μPAD) for the determination of copper in fresh and marine waters. A functionalized rhodamine-based chelator was synthesized and used as a chromogenic reagent, forming a highly intense pink complex with the analyte. The aim was to create a paper device that offers optimal performance and provides in-situ, rapid and cost-effective analysis in line with World Health Organization guidelines. The influence on the determination of several physical and chemical parameters was evaluated aiming to achieve the best performance. Under optimised conditions, a linear correlation was established in the range of 0.05–0.50 mg L−1 of copper, with a limit of detection of 10 μg L−1. The accuracy of the proposed method was assessed by comparing the results obtained with the developed μPAD and the results obtained with Inductively Coupled Plasma measurements (RE < 10 %). Recovery studies were also performed using different types of water samples with no need for any prior sample pre-treatment: tap, well, river and seawater. The average recovery percentage of 101 % (RSD = 4.3 %) was obtained, a clear indication of no multiplicative matrix interferences.pt_PT
dc.description.versioninfo:eu-repo/semantics/publishedVersionpt_PT
dc.identifier.doi10.1016/j.talanta.2024.125683pt_PT
dc.identifier.eid85183625839
dc.identifier.issn0039-9140
dc.identifier.pmid38301372
dc.identifier.urihttp://hdl.handle.net/10400.14/43924
dc.identifier.wos001176853400001
dc.language.isoengpt_PT
dc.peerreviewedyespt_PT
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/pt_PT
dc.subjectCopper(II) determinationpt_PT
dc.subjectFunctionalized copper ligandpt_PT
dc.subjectIn-situ analysispt_PT
dc.subjectNatural waterspt_PT
dc.subjectPaper sensor devicept_PT
dc.titleUse of a rhodamine-based chelator in a microfluidic paper-based analytical device for the in-situ copper quantification in natural waterspt_PT
dc.typejournal article
dspace.entity.typePublication
oaire.citation.titleTalantapt_PT
oaire.citation.volume271pt_PT
rcaap.rightsopenAccesspt_PT
rcaap.typearticlept_PT

Files

Original bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
93366996.pdf
Size:
2.43 MB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
3.44 KB
Format:
Item-specific license agreed upon to submission
Description: