Logo do repositório
 
Publicação

On-hand tool for ammonium and urea determination in saliva to monitor chronic kidney disease – design of a couple of microfluidic paper-based devices

dc.contributor.authorFerreira, Francisca T. S. M.
dc.contributor.authorMesquita, Raquel B. R.
dc.contributor.authorRangel, António O. S. S.
dc.date.accessioned2023-08-11T12:12:16Z
dc.date.available2023-08-11T12:12:16Z
dc.date.issued2023-10
dc.description.abstractIn this work, two microfluidic paper-based analytical devices (μPADs) were developed for the quantification of urea and NHx in human saliva to aid in the diagnosis/monitoring of chronic kidney disease (CKD). The NHx determination was based on the conversion of ammonium to ammonia, followed by its diffusion through a hydrophobic membrane and then the color change of bromothymol blue (BTB) indicator. In the urea determination, prior to the ammonium conversion and BTB color change, the enzymatic conversion of urea into ammonium was produced, using urease. Several optimization studies were carried out to attain a quantification range of 0.10–5.0 mM with 0.032 mM limit of detection for the NHx μPAD, and a determination range of 0.16–5.0 mM with 0.049 mM limit of detection for the urea μPAD. The method accuracy was assessed, and the measurements obtained with NHx μPAD were compared with the ones obtained from an ammonia ion selective electrode; while the measurements of the urea μPAD were compared with the ones obtained from a commercially available kit. There were no statistically significant differences between methods, proving that both NHx and urea μPAD were effective on-hand tools for CKD monitoring in saliva. To evaluate their functionality as point-of-care devices, stability studies were also performed and revealed that both NHx and urea μPAD were stable when stored in a vacuum for 2 and 1 month, respectively. After the sample introduction, the NHx μPAD could be scanned within the first 2 h and the urea μPAD within 1 h.pt_PT
dc.description.versioninfo:eu-repo/semantics/publishedVersionpt_PT
dc.identifier.doi10.1016/j.microc.2023.109102pt_PT
dc.identifier.eid85166011624
dc.identifier.issn0026-265X
dc.identifier.urihttp://hdl.handle.net/10400.14/42069
dc.identifier.wos001050007600001
dc.language.isoengpt_PT
dc.peerreviewedyespt_PT
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/pt_PT
dc.subjectCKDpt_PT
dc.subjectSaliva samplept_PT
dc.subjectSensorspt_PT
dc.subjectUrease conversionpt_PT
dc.subjectμPADspt_PT
dc.titleOn-hand tool for ammonium and urea determination in saliva to monitor chronic kidney disease – design of a couple of microfluidic paper-based devicespt_PT
dc.typejournal article
dspace.entity.typePublication
oaire.citation.titleMicrochemical Journalpt_PT
oaire.citation.volume193pt_PT
rcaap.rightsopenAccesspt_PT
rcaap.typearticlept_PT

Ficheiros

Principais
A mostrar 1 - 1 de 1
Miniatura indisponível
Nome:
74051836.pdf
Tamanho:
1.08 MB
Formato:
Adobe Portable Document Format
Licença
A mostrar 1 - 1 de 1
Miniatura indisponível
Nome:
license.txt
Tamanho:
3.44 KB
Formato:
Item-specific license agreed upon to submission
Descrição: