Comparative studies of different signal acquisition systems in microfluidic paper-based analytical devices developed for accurate point-of-care analysis

dc.contributor.authorPereira, Mafalda G.
dc.contributor.authorBrandão, Teresa R. S.
dc.contributor.authorRangel, António O. S. S.
dc.contributor.authorMesquita, Raquel B. R.
dc.contributor.authorCerdà, Víctor
dc.date.accessioned2025-12-15T12:42:42Z
dc.date.available2025-12-15T12:42:42Z
dc.date.issued2026-01-01
dc.description.abstractThe exponential growth of small, on-hand devices/sensors has been aligned with the increasing use of smartphones as detectors. However, smartphone detection may result in high variability of signal acquisition, due to differences in lighting conditions, consequently compromising accurate and reliable quantification. The use of smartphones for qualitative assessment (“yes/no” response) has been widely employed; however, their use in quantitative assessment has been less reported. For point of care (POC) analysis, the use of smartphones as detectors can be an exceptional advantage; however, it is crucial to ensure that the accuracy of the analysis is not compromised. A critical and structured evaluation of image acquisition systems was performed to evaluate the impact on POC biomarkers quantification with microfluidic paper-based devices (?PADs). Different image acquisition systems were tested for a ?PAD to properly evaluate the impact of lighting conditions when using a smartphone in a specially designed box. The decrease in intensity counts was about 30% from the scanner to the smartphone, but it did not affect the calculated absorbance. We proved that smartphones can be used as reliable detection systems, providing accurate and precise quantification when appropriate attention is given to exterior lighting conditions. When quantifying iodide and nitrite in a sample, a relative error of 5% was observed for both image-acquiring systems, a scanner and a smartphone.eng
dc.identifier.citationPereira, M. G., Brandão, T. R. S., Rangel, A. O. S. S., & Mesquita, R. B. R. et al. (2026). Comparative studies of different signal acquisition systems in microfluidic paper-based analytical devices developed for accurate point-of-care analysis. Microchemical Journal, 220, Article 116337. https://doi.org/10.1016/j.microc.2025.116337
dc.identifier.doi10.1016/j.microc.2025.116337
dc.identifier.eid105022813989
dc.identifier.issn0026-265X
dc.identifier.otherbc2e1da4-9287-4050-ae1d-b085e9b270ce
dc.identifier.urihttp://hdl.handle.net/10400.14/55915
dc.language.isoeng
dc.peerreviewedyes
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/
dc.subjectAccurate quantification
dc.subjectSmartphone detection
dc.subjectUser-friendly analysis
dc.subject?PAD
dc.titleComparative studies of different signal acquisition systems in microfluidic paper-based analytical devices developed for accurate point-of-care analysiseng
dc.typeresearch article
dspace.entity.typePublication
oaire.citation.titleMicrochemical Journal
oaire.citation.volume220
oaire.versionhttp://purl.org/coar/version/c_970fb48d4fbd8a85

Ficheiros

Principais
A mostrar 1 - 1 de 1
A carregar...
Miniatura
Nome:
134941588.pdf
Tamanho:
3.74 MB
Formato:
Adobe Portable Document Format