Repository logo
 
No Thumbnail Available
Publication

Chloride pseudotitration in wines by FIA with a Ag2S/Ag tubular electrode as detector

Use this identifier to reference this record.
Name:Description:Size:Format: 
19505030.PDF2.21 MBAdobe PDF Download

Advisor(s)

Abstract(s)

The determination of chloride in wines by potentiometric titration with silver cation in a nonsegmented flow system (FIA) using a tubular electrode of silver/silver sulfide detector is described. There was good agreement between the results obtained using the FIA titration and those obtained using the reference procedure. The methodology developed is adequate for the determination of chloride in various types of wines with chloride contents between 5 and 600 mg/liter, at sampling rates of 60 and 120 samples per hour.

Description

Keywords

Citation

LIMA, José L. F. C.; RANGEL, António O. S. S - Chloride pseudotitration in wines by FIA with a Ag2S/Ag tubular electrode as detector. Journal of Food Composition and Analysis. ISSN 0889-1575. Vol. 2 n.º 4 (1989), p. 356-363