In this work, a sequential injection system for the pseudo-titration of chloride in milk is described. Milk was directly aspirated
into the system and sandwiched between two silver nitrate plugs (titrant). The aspirated zones were then propelled to the detector
(Ag2S/Ag tubular electrode), where the depletion in the titrant silver concentration (due to the formation of a AgCl precipitate) was
monitored. The results obtained by the developed sequential injection titration method were not statistically different from those
provided by the potentiometric reference procedure with relative standard deviations better than 3.4%. The sampling rates achieved
were 17 samples per hour.
Description
Keywords
Sequential injection Milk Chloride pseudo-titration
Pedagogical Context
Citation
LIMA, M. J. Reis ; FERNANDES, Sílvia M. V. ; RANGEL, António O. S. S. - Sequential injection titration of chloride in milk with potentiometric detection. Food Control. ISSN 0956-7135. Vol. 15, n.º 8 (2004), p. 609-613
Publisher