dc.description.abstractreactive phosphorus in water samples is described. The system includes a liquid waveguide long pathlength cell connected to a charge coupled device (CCD) spectrophotometer. Calibration was linear up to 500 µgPL−1, with a detection limit of 17 µgPL−1 and a quantification limit of 56 µgPL−1. An injection throughput of 60 determinations h−1 and repeatability (R.S.D.) of 2.2% were achieved. Potential interference from silicate was effectively masked by addition of a tartaric acid stream. The accuracy of the proposedmethodology was assessed through analysis of a certified reference material and recovery tests on water samples. The developed procedure allows the determination of phosphorus in water samples at trace levels with high sensitivity, reduced reagent consumption and low waste production.por
dc.description.versioninfo:eu-repo/semantics/publishedVersion
dc.identifier.citationNeves, M. Sofia A. C.; Souto, M. Renata S.; Tóth, lldikó V.; Victal, Sérgio M. A.; Drumond, M. Cláudia; Rangel, António O. S. S. - Spectrophotometric flow system using vanadomolybdophosphate detection chemistry and a liquid waveguide capillary cell for the determination of phosphate with improved sensitivity in surface and ground water samples – In International Conference on Flow Injection, 14th. Berlin, Germany, 3-7 September, 2007. "Talanta". ISSN 0039-9140. 77: 2 (2008) 527–532por
dc.identifier.doi10.1016/j.talanta.2008.03.014
dc.identifier.urihttp://hdl.handle.net/10400.14/2761
dc.language.isoengpor
dc.publisherElsevierpor
dc.subjectPhosphate determinationpor
dc.subjectFlow injectionpor
dc.subjectLiquid waveguide capillary cellpor
dc.subjectLong pathlength spectrophotometrypor
dc.subjectSilicate interferencepor
dc.titleSpectrophotometric flow system using vanadomolybdophosphate detection chemistry and a liquid waveguide capillary cell for the determination of phosphate with improved sensitivity in surface and ground water samplespor
dc.typejournal article
dspace.entity.typePublication
person.familyNameTóth
person.familyNameSantos Silva Rangel
person.givenNameIldikó
person.givenNameAntónio Osmaro
person.identifier.ciencia-idD61D-7CFB-178A
person.identifier.orcid0000-0002-7342-1063
person.identifier.orcid0000-0002-6486-8947
person.identifier.ridG-1733-2010
person.identifier.scopus-author-id35512332800
person.identifier.scopus-author-id7004827387
rcaap.rightsrestrictedAccesspor
rcaap.typearticlepor
relation.isAuthorOfPublication7889563f-f3fc-4057-a97c-b51425b4dc46
relation.isAuthorOfPublication467be022-fd9c-4549-aa48-48ebcc1098eb
relation.isAuthorOfPublication.latestForDiscovery467be022-fd9c-4549-aa48-48ebcc1098eb

Ficheiros

Principais
A mostrar 1 - 1 de 1
Miniatura indisponível
Nome:
1-s2.0-S0039914008001951-main.pdf
Tamanho:
223.3 KB
Formato:
Adobe Portable Document Format
Licença
A mostrar 1 - 1 de 1
Miniatura indisponível
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição: