Rangel, António O. S. S.Tóth, Ildikó V.2011-10-142011-10-141996RANGEL, António O. S. S. ; TÓTH, Ildikó V. - Flow injection sequential speciation of free and total potassium in fortified wines. Analytical Sciences. ISSN 0910-6340. 12 (1996) 887-891http://hdl.handle.net/10400.14/6091A flow injection system for the potassium speciation in fortified wines, using two detectors placed in series, is described. The free, or ionized form, is monitored in a tubular-shaped potassium ion-selective electrode, and the total potassium content in a flame emission spectrometer used as end-of-line detector. Potassium determinations can be carried out in a linear range from 300 to 2000 mg 1-1, with a sampling-rate of 60 samples per hour. The results obtained were comparable to the conventional procedures using the same detection systems, and the precision was good (relative standard deviations below 3.1%o for free potassium and 3.8% for total potassium).engFlow injectionPotassium speciationWinePotentiometryFlame emission spectrometryFlow injection sequential speciation of free and total potassium in fortified winesjournal article10.2116/analsci.12.887