Utilize este identificador para referenciar este registo: http://hdl.handle.net/10400.14/6865
Título: Chloride pseudotitration in wines by FIA with a Ag2S/Ag tubular electrode as detector
Autor: Lima, Jose L. F. C.
Rangel, António O. S. S.
Data: 1989
Editora: Elsevier
Citação: LIMA, José L. F. C.; RANGEL, António O. S. S - Chloride pseudotitration in wines by FIA with a Ag2S/Ag tubular electrode as detector. Journal of Food Composition and Analysis. ISSN 0889-1575. Vol. 2 n.º 4 (1989), p. 356-363
Resumo: The determination of chloride in wines by potentiometric titration with silver cation in a nonsegmented flow system (FIA) using a tubular electrode of silver/silver sulfide detector is described. There was good agreement between the results obtained using the FIA titration and those obtained using the reference procedure. The methodology developed is adequate for the determination of chloride in various types of wines with chloride contents between 5 and 600 mg/liter, at sampling rates of 60 and 120 samples per hour.
Peer review: yes
URI: http://hdl.handle.net/10400.14/6865
Versão do Editor: doi:10.1016/0889-1575(89)90007-0
Aparece nas colecções:ESB - Artigos em revistas internacionais com Arbitragem / Papers in international journals with Peer-review

Ficheiros deste registo:
Ficheiro Descrição TamanhoFormato 
Chloride pseudotitration in wines by FIA with....PDF2,26 MBAdobe PDFVer/Abrir


FacebookTwitterDeliciousLinkedInDiggGoogle BookmarksMySpace